Protecting healthcare professionals and patients with gentle light

Equipped with high color rendering LEDs that approach the properties of natural light
Patients are nervous before an operation. In order to calm them and help them feel at ease.
Our design gives them a soft impression.

Recommended UsesOperating room/Emergency Room/Delivery Room

"Ultra-high color rendering LEDs" achieve reliable visibility and distinguishability

We used “Ultra-high color rendering LEDs” to achieve a high index of 95 for the R9 value. This corresponds to red(blood) in the special color rendering index (Ri), an index that includes colors close to the human body. This makes the colors of blood,organs, and tissue in the operative field more visible and distinguishable, helping improve the precision of operations.

Light that is gentle on your eyes with reduced risk of blue light

Doctors who continue to look at an operation site under an abnormally bright light experience tired eyes due to blue light. “Our LED” reduce the blue light risk and achieve a spectrum that does not strain the eyes.

Low-glare light that makes the operation site visible and improves operation efficiency

The unique optical design prevents glare when illuminated on the operation site. This significantly reduces eye strain and discomfort for doctors, who must continue to look at the operation site over a long period of time.

Light with "zero flickering" reduces strain on your eyes

Commonly-used lights adjust their brightness by repeatedly turning on and off in an extremely short interval (AC lighting, PWM control). However, the minute “flickering” strains your eyes, even if you’re not aware of it. This also applies to the light environment of the operation room. With our DC lighting, the light can be constantly “ON”, providing a light that does not flicker and is easy on your eyes.

Illuminating with soft edge of lgith field

A strong light gives a surgeon’s eyes stress due to a brightness when a surgical field is illuminated by operation light. This strong light makes a contrast for an edge of light fields.

Turning on/off a light gradually

Sudden lighting makes eye blind. However, Gentle lighting on/off system of cell reduces
eye strain.

Heat dissipation via handle

Excess heat, which can affect internal electronics, is released through the aluminum hand grip
It helps heat dissipate and reduces an affection for electronics
*It does not become hot

Specifications

Model Number DC10
General Name(PMDA) Surgical lighting equipment
Composition Wall control panel + Support arms + Lighthead
Lighthead diameter 670mm
Light source Ultra high color rendering LED
Illumination method Reflector method
Number of LED units Number of units / LEDs : 12
Center illuminance at 1m(irradiance) About 130,000lux(529W/m²)
Maximum illuminance(irradiance) About 160,000lux(650W/m²)
Light field diameter(d10) About 215mm / 300mm
Color temperature About 4,250K
Color rendering index Ra : 99 R9 : 95
Illuminance adjustment 10 to 100%(8steps) Approx. 1%(ENDOSCOPE mode)
by wall control panel / Operation control switch
Focus adjustment 850 to 1,500mm
by sterilizable center handle / Focus knob(2 locations)
LED expected life 40,000hours(time to 70% illumination)*1*2
Recommended operating environment Temperature : 5 to 35°, Humidity : 30 to 70%, Air pressure : 800 to 1,060hPa
Input rated voltage AC100 – 240V 50/60Hz
Power consumption 90-100VA
Fuse rating 2A

*1:It is not a guaranteed value because it depends on the use status and conditions

*2:When the initial illuminance is lower than 70%. It is the life of the LED used and not the life of the product

*Please note in advance that partial changes in the design and specifications may be made without prior notice for reasons such as improvements

*Comply with IEC60601-1, IEC60601-1-2 and IEC60601-2-41

  • Consultations & Inquiries
  • Reserve an Observation of Actual Equipment
  • Download the Catalog
  • Support Information